Patient Monitor User Manual Product Specification
- 282 -
☆
Alarm
Providing alarms of EtCO
2
, FiCO
2
, EtN
2
O , FiN
2
O , EtAA, FiAA,
AwRR
Support:
Real-time gas concentration monitoring
Zero calibration
Interfering Gas Effects:
Gas Gas
Level
CO
2
Agents
N
2
O
IRMA AX+
N
2
O
4)
60
vol%
_ 1&2)
_ 1)
_ 1)
HAL
4)
4
vol%
_ 1)
_ 1)
_ 1)
ENF, ISO, SEV
4)
5
vol%
_ 1)
_ 1)
_ 1)
DES
4)
15
vol%
_ 1)
_ 1)
_ 1)
Xe (Xenon)
4)
80 vol%
-10% of reading
3)
_ 1)
_ 1)
He (Helium)
4)
50 vol%
-6% of reading
3)
_ 1)
_ 1)
Metered Dose Inhaler
Propellants
4)
Not for use with metered dose inhaler propellants
C
2
H
5
OH (Ethanol)
4)
0.3
vol%
_ 1)
_ 1)
_ 1)
C
3
H
7
OH
(Isopropanol)
4)
0.5 vol%
_ 1)
_ 1)
_ 1)
CH
3
COCH
3
(Acetone)
4)
1 vol%
_ 1)
_ 1)
_ 1)
CH
4
(Methane)
4)
3
vol%
_ 1)
_ 1)
_ 1)
CO
(Carbon monoxide)
5)
1 vol%
_ 1)
_ 1)
_ 1)
NO
(Nitrogen monoxide)
5)
0.02 vol%
_ 1)
_ 1)
_ 1)
O
2
5)
100
vol%
_ 1&2)
_ 1)
_ 1)
Note 1
: Negligible interference, effect included in the specification “Accuracy, all conditions”
above.
Note 2
: For probes not measuring N
2
O and/or O
2
the concentrations shall be set from host
according to the instructions. (IRMA AX+ does not measure O
2
)