Patient Monitor User Manual Monitoring CO
2
- 154 -
WARNING
32
Too strong positive or negative pressure in the patient circuit might affect the sample
flow.
33
Strong scavenging suction pressure might affect the sample flow.
34
To avoid electric shock, always physically disconnect the module and all patient
connections before cleaning.
35
Do not attempt to remanufacture, recondition or recycle the module as these
processes may damage the electrical components, potentially leading to patient
harm.
CAUTION
1
Masimo sidestream CO
2
module does not contain any user-serviceable parts.
2
Anyone not properly trained and certified must not attempt to perform any service or
maintenance of the Masimo product.
3
Only use existing mounting holes. Drilling additional holes will void the warranty.
4
Do not operate the module outside of the specific operating environment.
5
The module should be mounted securely to avoid risk of damage.
NOTE:
Use and store the module in accordance with specifications. See the Specifications
section in this manual.
NomoLine Sampling Lines
Masimo Sidestream CO
2
module samples gas from the respiratory circuit through the NomoLine
Family sampling line at a rate of 50 sml/min, making measurements of CO
2
possible for adult,
pediatric, infant and neonatal patients. The NomoLine Family of sampling lines are designed for
optimal performance and measurement fidelity when used with the Masimo Sidestream CO
2
module.
NomoLine sampling lines include nasal and nasal/oral cannulas for non-intubated patients with
and without supplementary oxygen delivery and airway adapter sets for intubated patients.