Patient Monitor User Manual Product Specification
- 255 -
A.7 SpO
2
Complies with ISO 80601-2-61: 2017.
EDAN Module
Measuring Range
0% to 100%
Resolution 1%
☆
Data Update Period
1 s
☆
Accuracy
☆
Adult /Pediatric
±
2% (70% to 100% SpO
2
)
Undefined (0% to 69% SpO
2
)
☆
Neonate
±
3% (70% to 100% SpO
2
)
Undefined (0% to 69% SpO
2
)
Sensor
Red Light
(660
±
3) nm
Infrared Light
(905
±
10) nm
Emitted Light Energy
< 15 mW
PI
Measuring Range
0-10, invalid PI value is 0.
Resolution 1
Nellcor Module
Measuring
Range
1% to 100%
Resolution 1%
☆
Data Update
Period
1 s
☆
Accuracy
DS-100A, OXI-A/N(Adult)
D-YS (Adult and Pediatric)
OXI-P/I (Pediatric)
±
3% (70% to 100% SpO
2
)
MAX-A, MAX-AL, MAX-N,
MAX-P, MAX-I, MAX-FAST
(Adult and Pediatric)
±2% (70% to 100% SpO
2
)