Patient Monitor User Manual Product Specification
- 247 -
Resolution 1
pause/min
ST value
Range
-2.0 mV to +2.0 mV
Accuracy
-0.8 mV to +0.8 mV: ±0.02 mV or 10%, whichever is greater.
Beyond this range: not specified.
Resolution 0.01
mV
QT measurement
Range
200 ms ~ 800 ms
Resolution 4
ms
Accuracy
± 30 ms
QTc measurement
Range
200ms ~ 800 ms
Resolution 1
ms
Δ
QTc measurement
Range
-600 ms ~ 600 ms
Resolution 1
ms
HR Averaging Method
Method 1
Heart rate is computed by excluding the minimum and maximum
values from the 12 most recent RR intervals and averaging the
residual 10 RR intervals.
Method 2
If each of three consecutive RR intervals is greater than 1200 ms,
then the four most recent RR intervals are averaged to compute
the HR.
Range of Sinus and SV Rhythm
Tachy
Adult: RR interval for 5 consecutive QRS complex
≤
0.5 s.
Pediatric/neonatal: RR interval for 5 consecutive QRS complex
≤
0.375 s.
Normal
Adult: 0.5 s < RR interval for 5 consecutive QRS complex < 1.5 s.
Pediatric/neonatal: 0.375 s < RR interval for 5 consecutive QRS
complex < 1 s.
Brady
Adult: RR interval for 5 consecutive QRS complex
≥
1.5 s.
Pediatric/neonatal: RR interval for 5 consecutive QRS complex
≥
1 s.