Patient Monitor User Manual EMC Information
- 290 -
Table 1 Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless
communications equipment
Test
frequency
(MHz)
Band
a)
(MHz)
Service
a)
Modulation
b)
Maximum
power (W)
Distance
(m)
Immunity
test level
(V/m)
385 380-390
TETRA
400
Pulse
modulation
b)
18 Hz
1.8 0.3 27
450 430-470
GMRS 460,
FRS 460
FM
c)
± 5 kHz
deviation
1 kHz sine
2 0.3 28
710
704-787
LTE Band
13, 17
Pulse
modulation
b)
217 Hz
0.2 0.3 9
745
780
810
800-960
GSM
800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation
b)
18 Hz
2 0.3 28
870
930
1720
1700-1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT; LTE
Band 1, 3, 4,
25; UMTS
Pulse
modulation
b)
217 Hz
2 0.3 28
1845
1970
2450 2400-2570
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse
modulation
b)
217 Hz
2 0.3 28
5240
5100-5800
WLAN
802.11 a/n
Pulse
modulation
b)
217 Hz
0.2 0.3 9
5500
5785
NOTE
If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the
transmitting antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1
m test distance is permitted by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50 % duty cycle square wave signal.
c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because
while it does not represent actual modulation, it would be worst case.