Patient Monitor User Manual Monitoring ECG
- 96 -
The picture below shows the conventional 10 electrodes placement.
Take the American standard for example; the 10 electrodes should be placed as follows:
Limb electrodes:
■
RA and LA: Place arm electrodes on the inside of each arm, between the wrist and the
elbow.
■
RL and LL: Place leg electrodes inside of each calf, between the knee and the ankle.
Chest electrodes:
■
V1: On the 4th intercostal space at the right sterna margin.
■
V2: On the 4th intercostal space at the left sterna margin.
■
V3: Midway between V2 and V4 electrodes.
■
V4: On the 5th intercostal space at the left clavicular line.
■
V5: On the left anterior axillary line, horizontal with V4 electrode.
■
V6: On the left middle axillary line, horizontal with V4 electrode.
10 Electrodes Placement