Patient Monitor User Manual Product Specification
- 271 -
gas
identification
concentration
2
nd
gas identification concentration
3
> 0.3vol%
When the concentration of DES in the
mixed gas is greater than 3%, the
identification concentration of the
second anesthetic gas should be
greater than 10% of the DES
concentration.
Sample Gas
Flowrate
150 ml/min, accuracy
±
15 ml/min
Warm-up
Time
Display reading within 20 s; reach to the designed accuracy within 2 minutes.
Rise Time
CO
2
/ N
2
O/ DES/ SEV/
ISO/ ENF
< 400ms
with 2 m gas sampling tube,
sample gas flowrate: 150 ml/min
HAL/ O
2
< 500 ms
Response
Time
< 4 s (with 2 m gas sampling tube, sample gas flowrate: 150 ml/min)
Work Mode
Standby, measure
O
2
Compensation
Manual: without O
2
module
Range: 0% to 100%
Resolution: 1%
Default: 16%
Auto: with O
2
module
Humidity
Compensation
Method
ATPD (default), BTPS
Zero
Calibration
Auto, Manual
☆
Alarm
EtCO
2
, FiCO
2
, AwRR, EtAA, FiAA, EtO
2
, FiO
2,
EtN
2
O, FiN
2
O
☆
Apnea
Alarm Delay
20 s, 25 s, 30 s, 35 s, 40 s; default value is 20 s.
Data Sample
Rate
100 Hz
Drift of
Measure
Accuracy
Meets the requirements of the measure accuracy
Barometric
Pressure
Compensation
Automatic (The change of barometric pressure will not add additional errors
to the measurement values.)
Note 1: This function is available in O
2
module.
Note 2: This function is available in G7, G7+ modules.
Note 3: This function is available in G7A, G7A+, G7S, and G7S+ modules.
Note 4: This function is available in G7S and G7S+ modules.