Patient Monitor User Manual Networked Monitoring
- 36 -
When it is set to
On
, the monitor can enter the normal working interface after start-up; when
it is set to
Off
, after start-up, a password window will be displayed, and the monitor can enter
the normal working interface until correct password is input. Default setting is
On
.
•
Select the minutes in
User login Timeout
. If there are no any operations to the monitor for
XX minutes (
5
,
15
,
30
,
60
and
Never
), the screen will enter into the screensaver status. User
Maintain password should be correctly input before user operates the monitor again. The
selection
Never
means the monitor will never enter into screensaver status and still in the
normal working status. Default setting is
Never
.
•
Set
Firewall
to
On
to protect against hacker attacking.
•
Set
Packets Limit
value for traffic monitoring. If the data traffic per minute exceeds the
preset threshold, the monitor will trigger the alarm “
Network Traffic Abnormity
” to remind
the user, and at the same time, the network will disconnect for 5 minutes. After 5 minutes,
the network will be re-connected and alarm disappears.
•
Set
CMS/Gateway Encryption
to
Off
,
TLS
or
AES
(default) when user connects the
monitor with network server (MFM-CMS/Gateway).
•
Set
HL7
to
On
/
Off
. The monitor supports HL7 protocol to upload data. To avoid hacker
attacking, setting HL7 to
Off
is normally recommended.
User can also set
HL7 IP
address of client-side in
User Maintain
>
Network Maintain
.
•
Set
HL7 Encryption
method to
Off
or
TLS
(default).
•
Click
Import Certificate
to install/upgrade the
Certificate
via USB flash drive. The
certificate issued by Certificate Authority (CA) is recommended and self-signed certificate
should be avoided. For detailed steps of importing certificates, please refers to service
manual.
NOTE:
1
When the monitor is turned on for the first time, modify the
User Maintain
password
according to the prompts. The default initial
User Maintain
password is
ABC
. After
modifying the password, please keep it safe.
2
When any password is continuously input incorrectly for more than 5 times, the
monitor will display the information:
More than five consecutive password errors.
4.2 Connecting the Wireless Network
Wi-Fi modules are optional to be configured in the monitors. And you should configure the
settings on the monitor following the steps below before connecting the monitor to a wireless
network:
1.
Select
Menu
>
Maintenance
>
User Maintain
, and input the password.
2.
In the
User Maintain
menu, select
Network Maintain
.
3.
In the
Network Maintain
menu, select
Wi-Fi
from the
Network Type
list. And click
Config
to open the
Wi-Fi Setup
window. The available networks will be listed in this
window.