Patient Monitor User Manual Monitoring NIBP
- 129 -
coming from the artery.
Patients on an edematous extremity.
13.4 Measurement Methods
There are four methods of measuring NIBP:
Manual
- measurement on demand.
Auto
- continually repeated measurements (between 1 and 480 minute adjustable interval).
The interval can be user defined, and the default interval of user defined is 2.5 minutes. After
the first measurement starts manually, the monitor will automatically measure NIBP as preset
interval. When the measurement interval is set to 5 min, 10 min, 15 min, 30 min, 60 min, 90
min, 120 min, 180 min, 240 min, 360 min and 480 min, the system will automatically adjust
the next measurement time. Here’s an example.
Auto Measurement Interval
Current Time
Next Measurement Time
5 min
12:02
12:05, 12:10, 12:15, 12:20, and so forth.
10 min
12:02
12:10, 12:20, 12:30, 12:40, and so forth.
15 min
12:02
12:15, 12:30, 12: 45, 13:00, and so forth.
30 min
12:02
12:30, 13:00, 13:30, 14:00, and so forth.
60 min
12:02
13:00, 14:00, 15:00, 16:00, and so forth.
90 min
12:02
13:00, 14:30, 16:00, 17:30, and so forth.
120 min
12:02
13:00, 15:00, 17:00, 19:00, and so forth.
180 min
12:02
13:00, 16:00, 19:00, 22:00, and so forth.
240 min
12:02
13:00, 17:00, 21:00, 1:00, and so forth.
360 min
12:02
13:00, 19:00, 1:00, 7:00, and so forth.
480 min
12:02
13:00, 21:00, 5:00, 13:00, and so forth.
When the completion time of manual measurement to the first hourly time is less than or equal to
30 seconds, the measurement will not be performed at the first hourly time, and the first
automatic measurement will be delayed to the next hourly time.
Continuous
- the measurement will run consecutively in five minutes, then the monitor
enters manual mode.
Sequence
- after the first measurement starts manually, NIBP measurements run
automatically according to the preset phase and interval. The phase can be selected as
4
,
5
and
6
. The interval can be set as
1 min
,
2 min
,
3 min
,
4 min
,
5 min
,
10 min
,
15 min
,
30 min
,
60 min
,
90 min
,
120 min
,
180 min
,
240 min
,
360 min
, and
480 min
. The user can also set
the measurement times in each phase, there are several selections:
Off
,
1
,
2
,
3
,
4
,
5
,
6
,
7
,
8
,
9
,
10
,
11
,
12
,
13
,
14
,
15
,
16
,
17
,
18
,
19
,
20
,
21
,
22
,
23
,
24
,
25
,
26
,
27
,
28
,
29
,
30
and
Continuous
.