Patient Monitor User Manual Networked Monitoring
- 39 -
2.
The real-time monitoring information is displayed on the central monitoring system as the
same to the monitor, and the central monitoring system can perform some bilateral control.
For example: changing patient information, receiving patient, discharging patient and so
forth.
For detailed information, please refer to
MFM-CMS Central Monitoring System User Manual
.
And the monitor supports HL 7 protocol.
NOTE:
1 Use wired instead of wireless networking when connecting the monitor to central
monitoring system in the operating room because the ESU will interfere with a
wireless network, which may cause networking failure.
2 Make sure the network connection between the monitor and MFM-CMS is in good
condition when the time synchronization function on the monitor is active. (Default
setting is on. Setting path:
Menu
>
Maintenance
>
User Maintain
>
Date/Time
Setup
>
Sync Time
). If the setting is on, the monitor will accept time synchronization
from MFM-CMS.
3 The time synchronization function might not be available to all software versions of
MFM-CMS. Consult our technical service department or your local distributor for
more information.
4 When deploying the network of the monitor and MFM-CMS, it is recommended to
isolate the network and the Intranet system of the hospital by using VLAN so as to
ensure the network security. Only trusted devices are allowed to join the VLAN
network.
4.5 Connecting the Monitor to CMS-LITE
The monitor can be connected to the CMS-LITE Data Management Software. Through the
network:
1.
The monitor sends patient information, monitoring or measurement data to the CMS-LITE
Data Management Software.
2.
The real-time monitoring information is displayed on the CMS-LITE Data Management
Software as the same to the monitor.
For detailed information, please refer to
CMS-LITE Data Management Software User Manual
.
4.6 Connecting the Monitor to Gateway
The monitor can be connected to the Gateway (V1.1 and above), which provides clinicians with
the capability of viewing and collecting patient data remotely and the data exchange of selected
clinical and administrative information between iM series Network and the hospital network.
To set the monitor server IP address, select
User Maintain
>
Network
>
Server IP
. Make sure
the monitor share the same server IP with the computer in which the MFM-CMS software is
installed.
For more information about Gateway communication, refer to
GW1
Gateway User Manual
.