Patient Monitor User Manual Other Functions
- 209 -
The monitor can detect the storage space threshold. Select
Menu
>
Common Function
>
Data
Store
and set
Threshold Detection
to
On
. When the removable device is newly inserted and
selected as storage device, and its remaining storage space is less than 300 M, the monitor will
not store data, prompting
The space in U disk is less than 300 M. Please clean it up.
The user
needs to clean up the space manually till the remaining space is more than 300 M, thus the
monitor will keep storing data. When this removable device is read-only and its space is
insufficient, the monitor only provides alarm of
Read-only storage device
.
NOTE:
1 The storage time varies according to the patient's parameter data volume. When the
single patient data store reaches 240 hours, the monitor will automatically create a
new folder for continuous data store. If you chose
Keep Storing
, when the storage
space is insufficient, the earliest folder will be deleted and new folder will be created.
2
Threshold Detection
is only applicable to the removable devices.
3 Without use of data store function, all data measured (including trend data, review
data, alarm events and so on) are cleared either when the monitor is turned off or
when the monitor is powered down in the process of monitoring.
26.3.2 Activating/ Deactivating Data Storing
To activate/ deactivate the data storing function, select
Menu
>
Maintenance
>
User
Maintain
>
Other Setups
, and set
Data Store
to
On
or
Off
.
The monitor will stop storing data in the storage device under the following circumstances:
No storage devices are selected.
There is no enough space in the storage device for storing data.
The removable device is read-only.
The storage device is damaged.
The data storing function is deactivated.
The monitor is switched off.
The power supply is off.
26.3.3 Selecting a Storage Device
To configure the storage device, select
Menu
>
Common Function
>
Data Store
. The initial
default storage device is
Internal Storage
. When the monitor has no internal storage device, the
storage device displays
null
.
When user switches the storage device from an internal device to a removable device or switches
from one removable device to another removable device, the user password is required.
After you configure the appropriate storage device, click exit. If the storage device is successfully
starting data storing, the monitor will be indicated by the symbol
. If there is no enough
space in storage device, or the storage device is read-only/damaged, the symbol
will be
displayed.