Patient Monitor User Manual Review
- 189 -
browse more ARR alarm events. You may select an alarm event and access the alarm review
interface to get more information. On the alarm review interface, you can:
Right or left shift the waveform to review the complete 8-second waveform.
Select
Record
and output the arrhythmia waveform by the recorder.
According to the actual clinical needs, select another name from the pull-down list of
Rename
for the arrhythmia event. Confirm the changes to make the settings take effect.
Select
Delete
to remove a specific arrhythmia event.
Select
Alarm List
or
exit to get back to the arrhythmia review interface.
NOTE:
1 If there are more than 200 arrhythmia events, the monitor will only keep the recent
ones.
2 The name of arrhythmia event will be shown on the alarm status area.
3 The renaming is only available for the ARR alarm event of the current patient, not for
that of the history patient.
21.6 12-lead Analysis Review
To review the 12-lead analysis result, please press the
Analysis Review
key
on the
screen or select
Menu
>
Review
>
Analysis Review
.
In the 12-lead analysis review window:
■
The user can switch between results and waveforms. Select
Waveform
to review the
analysis waveforms and
Result
to review the analysis results.
■
Select
Delete
to delete the analysis results displayed on the current screen.
■
Select
and
to browse more analysis results or waveforms.
■
Select
Record
to print out the analysis results by the recorder.
■
Select
to print out the analysis report by the printer.
21.7 ST Segment Review
To review the ST segment, please press the
ECG Setup
>
ST Analysis
>
ST Segment Review
.
In the ST segment review window,
The user can select the lead waveform that want to review.
The user can select the ST segment to review. There are 20 groups of segments at the most,
user can review one ST segment, and can also review all overlapped ST segments.
The color of ST waveform is consistent with the color of ECG. When only one ST segment
is reviewed, this segment is highlighted, the ST value and saved time of the ST Segment is
displayed, at the same time, the color of other segments becomes dark.