Patient Monitor User Manual Monitoring NIBP
- 131 -
6.
Wait until the first reading is taken.
NOTE:
1 The width of the cuff is either approximately 40% of the limb circumference or 2/3 of
the upper arm length. The inflatable part of the cuff should be long enough to encircle
80%-100% of the limb. The wrong size of cuff can cause erroneous readings. If the
cuff size is in question, use another cuff with suitable size to avoid errors.
2 If an NIBP measurement is suspect, repeat the measurement. If you are still
uncertain about the reading, use another method to measure the blood pressure.
3 Please make sure the cuff is well connected. A leak of air may cause measurement
error.
4 Please select the cuff with the suitable size. An unsuitable cuff may cause incorrect
measurements.
5 Avoid incursion of liquid into the cuff. If this happens, please desiccate the cuff
completely.
13.5.1 Operation Prompts
1.
Manual Measuring
Access the
NIBP Setup
menu and set the
Measure Mode
item to
Manual
. Then press the
button on the front panel or shortcut key
on the main screen to start a manual
measurement.
2.
Automatical Measurement
Access the
NIBP Setup
menu and set the
Measure Mode
item to
Auto
, select time interval as need,
then press the
button on the front panel or shortcut key
on the main screen.
3.
Continuous measurement
Access the
NIBP Setup
menu and pick the
Continuous
item to start a continuous measurement.
The continuous measurement will last 5 minutes.
4.
Sequence measurement
Access the
NIBP Setup
menu and set the
Measure Mode
item to
Sequence
, then click
Sequence
.
In
Sequence Measurement Setup
, set the
Phase Counts
,
Times
and
Interval
to start a sequence
measurement.
5.
Stopping continuous measurement
During continuous measurement, press the
button on the front panel or shortcut key
on the main screen at any time to stop measurement.
13.5.2 Correcting the Measurement if Limb is not at Heart Level
To correct the measurement if the limb is not at heart level to the displayed value: