Patient Monitor User Manual Maintenance
- 222 -
Maintenance and Test Schedule
Frequency
Safety checks. Selected tests on
the basis of IEC60601-1
At least once every two years, or as needed, after any
repairs where the power supply is removed or replaced, or
if the monitor has been dropped.
Check all monitoring functions
and measuring functions
At least once every two years, or as needed.