Patient Monitor User Manual Monitoring NIBP
- 134 -
If
Manual
is chosen, the preset value by users will be adopted as the inflation value when
measuring blood pressure.
If
AUTO
is chosen, the default value will be adopted as the inflation value when
measuring blood pressure.
13.11 Cleaning Mode
The cleaning mode can remove the dust and foreign matters in the air valve to ensure the
accuracy of NIBP measurement. To start the cleaning mode, please select
User Maintain
>
NIBP Maintain
>
Cleaning Mode
, the monitor displays:
Be sure the cuff is disconnected from
monitor
, after confirmation and clicking
Start Cleaning
button, cleaning mode starts. The
cleaning mode lasts three minutes. In this mode, the monitor displays
Cleaning in progress
,
the
remaining time of cleaning mode and
cuff value are also displayed. When the countdown finishes,
the monitor exits cleaning mode automatically, if the user needs to exit the cleaning mode in
advance, please click
Stop
button.
When the air pressure is abnormal, the monitor will automatically turn off the cleaning mode and
display the prompt message:
Cleaning failed
.
NOTE:
Cleaning mode is only available when the patient type is adult.
13.12 Assisting Venipuncture
The user can use the NIBP cuff to cause a pressure close to diastolic pressure, so as to block the
venous blood vessel and therefore help venipuncture. To assist venipuncture:
1.
Select
NIBP Setup
>
Venipuncture
;
2.
Select the appropriate
Cuff Pressure
according to the patient type;
3.
Select
Start
, the monitor displays:
Venipuncture Starting
;
4.
Wait until the monitor prompts
I
n venipuncture process
. If an abnormal alarm occurs
before it, no follow-up operation can be carried out. Restart the procedure after checking if
necessary;
5.
Puncture vein and draw blood sample;
6.
Select
Stop
to deflate the cuff. If you do not deflate the cuff, the cuff automatically deflates
when the venipuncture time expires (170 seconds for adult and pediatric patient, 85 seconds
for neonatal patient).
During venipuncture, pay attention to the cuff pressure and the countdown displayed in the NIBP
numerics area. When the remaining time is 30 seconds, the monitor issues a reminder tone and
the countdown displays in red, prompting the user that the venipuncture time is to expire.
NOTE:
1 Only when the monitor exits
Venipuncture
menu, the user can do other operations.
2 When the monitor is in DEMO mode, continuous measurement process, manual
measurement process, sequence measurement process or auto measurement
process, Assisting Venipuncture function is not available.