Patient Monitor User Manual Product Specification
- 253 -
Alarm Type
SYS, DIA, MAP
Cuff pressure Measuring Range
0 mmHg to 300 mmHg
Pressure Resolution
1 mmHg
Measuring Accuracy
☆
Maximum Mean
Error
±5 mmHg
☆
Maximum Standard
Deviation
8 mmHg
Adult/Pediatric Neonate
Normal
Condition
Single Fault
Condition
Normal
Condition
Single fault
Condition
Maximum Cuff
Pressure
300 mmHg
330 mmHg
150 mmHg
165 mmHg
Maximum Measuring
Period
Less than 160 s Less than 180 s
Less than 80 s
Less than 90
s
Pre-inflation Pressure
Adult/ Pediatric Mode
Default: 180 mmHg
Range: 120/140/150/160/180/200/220/240/260/280 mmHg
Neonatal Mode
Default: 120 mmHg
Range: 80/100/120/140 mmHg
Dual Independent Channel Overpressure Protection
Adult/Pediatric
< 300 mmHg
Neonatal
< 150 mmHg
SunTech Module
Not applicable to iM50 or M50.
Method Oscillometric
Mode
Manual, Auto, Continuous, Sequence
Measuring Interval in AUTO Mode
(unit: minute)
1/2/3/4/5/10/15/30/60/90/120/180/240/360/480 and
User Define
☆
Measuring Parameter
SYS, DIA, MAP, PR
Pressure Unit
kPa, mmHg, cmH
2
O
☆
Measuring Range