Patient Monitor User Manual Product Specification
- 270 -
Range N
2
O 0~100
vol%
Halothane/ Enflurane/
Isoflurane
0~8 vol%
Sevoflurane 0~10
vol%
Desflurane 0~22
vol%
O
2
1
0~100%
Resolution
N
2
O, O
2
1
1%
CO
2
, AG
0.1%
AwRR
Measurement range
2 ~ 150 rpm
Measuring accuracy
±
1 bpm (120 bpm and below),
Not specified (120 bpm above)
Resolution
1 rpm
☆
Accuracy
CO
2
±
(0.2vol% + 2% of
reading)
Typical conditions:
Ambient temperature: (25
±
3)
°
C
Barometric pressure: (760
±
10) mmHg
Balance gas: N
2
Sample gas flowrate: 100
ml/min
N
2
O
±
(2vol% + 2% of
reading)
Hal, Enf, Iso, Sev,
Des
±
(0.15vol% + 5% of
reading)
O
2
1
±
(1vol% + 2% of
reading)
CO
2
±
(0.3vol% + 4% of
reading)
Claimed working
environment
N
2
O
±
(2vol% + 5% of
reading)
Hal, Enf, Iso, Sev,
Des
±
(0.2vol% + 10% of
reading)
O
2
1
±
(2vol% + 2% of
reading)
Anesthetic
gas
identification
method
Manually set anesthetic gas
1
Automatic identification of an anesthetic gas
2
Automatic identification of two anesthetic gases
3
Anesthetic 1
st
gas identification concentration
2
>
0.2vol%