Patient Monitor User Manual Product Specification
- 250 -
QT/QTC (ms)
P-QRS-T AXIS
A.5 RESP
Method
Impedance between RA-LL, RA-LA
Measurement lead
Options are lead I and II. The default is lead II.
Calculation Type
Manual, Automatic
Baseline Impedance Range
200
Ω
to 2500
Ω
(with ECG cables of 1 K
Ω
resistance)
Measuring Sensitivity
Within the baseline impedance range: 0.3
Ω
Waveform Bandwidth
0.2 Hz to 2.5 Hz (-3 dB)
Respiration Excitation
Waveform
Sinusoid, 45.6 kHz (
±
10%), < 350
μ
A
☆
RR Measuring Range
☆
Adult
0 rpm to120 rpm
☆
Neo/Ped
0 rpm to150 rpm
Resolution 1
rpm
☆
Accuracy
☆
Adult
6 rpm to 120 rpm:
±
2 rpm
0 rpm to 5 rpm: not specified
☆
Neo/Ped
6 rpm to 150 rpm:
±
2 rpm
0 rpm to 5 rpm: not specified
☆
Gain Selection
×
0.25,
×
0.5,
×
1,
×
2,
×
3,
×
4,
×
5
☆
Sweep
6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s
☆
Apnea Alarm Time Setup
10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s; default value is 20 s.
A.6 NIBP
Complies with IEC 80601-2-30: 2009+A1: 2013.
EDAN Module
Technique Oscillometry