SEFELEC1500M Instructions
75
V1.06
9.4.
SETTING THE COMPARISON THRESHOLDS
The device offers two comparison thresholds Imin and Imax.
The main threshold for current measurement is the Imax threshold, if the current measured is
greater than the Imax threshold the result is FAIL, if the current measured is less than Imin the
result is PASS.
The Imin threshold makes it possible if needed to detect an incorrect connection on the tested
sample which would then have an abnormally low current value.
The value of Imax must always be higher than the value of Imin.
To access the settings for Imin and Imax scroll the measurement parameters screen to display: