SEFELEC1500M Instructions
21
V1.06
4.3.
PRECAUTIONS FOR USE
IMPORTANT
The measuring device must not be used to switch or handle external
sources of energy that are not controlled by its system software. These
power sources include relay solenoids, inductors without flyback diodes
and undischarged capacitors.
IMPORTANT
During insulation measurements on capacitance > 1 mF, setting up a
discharge system that makes it possible to ensure that the capacitor is
discharged is mandatory.
The DISCHARGE command makes it possible to check the discharge
level of the capacitor.
If this instruction is not complied with, the measuring device can be
damaged.
NOTE
As the SEFELEC1500M is a measuring instrument, EATON cannot be held
liable for any equipment faults or loss of production resulting from damage to
the tester by failure to comply with the operating conditions set out above (en-
vironment, mains power supply and operation).