SEFELEC1500M Instructions
146
V1.06
LLIM (LLIMit) <NR3>
Changes the value of the lower threshold of the current parameter set for the active
function.
E.g.:
MEG:LLIM 1.0E+6
(new value in Ohms)
PIC:LLIM 1.50E-3
(new value in Amperes)
The unit is automatically expressed in Volts, Ohms or Amperes and depends on the
function.
TIM (TIMe) AUT / MAN / FAIL
Selects the
AUT
timer modes for automatic,
MAN
for manual,
FAIL
for fault.
FILT (FILTer) NOR / SLO / FAS / CAP1 / CAP2 /
Activates the measuring modes
NORMAL, SLOW, FAST, CABLE and CAPACITOR
E.g.:
CONF:FILT NOR
CONF:FILT CAP1
RANGE <NR1>
with <NR1> between 0 and 10, 0 corresponding to AUTO range, refer to 7.6 and 9.6
section for current or resistance correspondence.
MKM ON
MKM OFF
Allows to activate or disactivate the Megohm.km mode
(Extended version « MEGOHMKM »)
MKMLEN <NR1>
with <NR1> cable length (1 to 16000 meters) in Megohm.km mode
(Extended version « MEGOHMKMLENGTH »)
MKMCOUNT <NR1>
with <NR1> number of wires (0 to 999) in Megohm.km mode
(Extended version « MEGOHMKMCOUNT »)
MEAS (MEASure)
Switches to measuring mode in the active function.
STOP
Interrupts the test in progress.
QUIT
Exits the active function.
SRQ
Through analogy with the Service Request command of the
IEEE488
bus, this code au-
thorises the sending of a
Z
character for reporting the events (end of test, format error,
safety circuit open...). This code must be sent at the beginning of the program after
switching to
REMOTE
mode.