SEFELEC1500M Instructions
119
V1.06
CTRLOUT (Output C7-5)
Validation contact for taking control of the device by the PLC.
EOT (Output C7-7)
End of test contact
PLCPASS (Output C7-8)
Test result contact: pass
PLCFAIL (Output C7-6)
Test result contact: fail
ERROR (Output C7-9)
Indicates if an error occurred during the test , the SR0 to SR3 outputs provide the error
type.
SR0 à SR3 (Output C8-5,6,7,8)
Error type coding :
SR3 SR2
SR1
SR0 Error type
0
0
0
0
SAFETY LOOP OPEN
0
0
0
1
UNDER RANGE
0
0
1
0
OVER RANGE
0
0
1
1
CHARGING
0
1
0
0
VOLTAGE ERROR
-
-
-
-
…
1
1
1
1
….
IN_OUT_0-10V_0 ( Output C6-1)
0-10VDC analog output - configurable
IN_OUT_0-10V_1 ( Output C6-2)
0-10VDC analog output - configurable
IN_0-10V_2 ( Input C8-9)
0-10VDC analog input - configurable
IN_ 0-10V_3 ( Input C7-3)
0-10VDC analog input
– configurable