7.3.
SETTING THE COMPARISON THRESHOLDS
The device offers two comparison thresholds Rmin and Rmax.
The main threshold for insulation resistance measurement is the Rmin threshold, if the re-
sistance measured is less than the Rmin threshold the result is FAIL, if the resistance measured
is greater than Rmin the result is PASS.
The Rmax threshold makes it possible if needed to detect an incorrect connection on the tested
sample which would then have an abnormally high insulation resistance value.
The value of Rmax must always be higher than the value of Rmin.
To access the settings for Rmin and Rmax scroll the insulation parameters screen to display: