SEFELEC1500M Instructions
154
V1.06
17.1.1.
SAFETY CONNECTOR C5
1
Safety loop circuit contact: to connect with 9
2
Safety loop circuit contact: to connect with 10
3
Green lamp
4
Red lamp
5
+25V common red-green lamps
6
0V
– ground
7
Contact PASS (GOOD) (= +25 V)
8
Contact FAIL (BAD) (= +25 V)
9
Safety loop circuit contact: to connect with 1
10
Safety loop circuit contact: to connect with 2
Connection example of the red/green lamps, of the contacts of the safety circuit and of
the pass/fail contacts
NOTE
The contacts of the safety circuit must be free of any potential (24 VDC, 230 VAC,
Ground, etc.) They are protected by thermal fuses, in the event of overcurrent wait
15 seconds before trying again.