SEFELEC1500M Instructions
56
V1.06
7.6.
SELECTING THE MEASUREMENT RANGE
The SEFELEC1500M has an automatic measurement range search mode.
This is the most practical operating mode for the user
. It is however possible in the case
where it is sought to increase the rapidity of the measurement to block the device in a range
(range change time: 150 msec. per range)
Press the active area of the "Range" parameters line:
Select the resistance range, the max. value indicated is equal to: Voltage / Imin and the min.
value is equal to Voltage / Imax.
With the following current ranges:
N°
Range
N°
Range
N°
Range
N°
Range
0
AUTO
3
20-200 µA
6
20-200 nA
9
20-200 pA
1
2-20 mA
4
2-20 µA
7
2-20 nA
10
2-20 pA
2
0.2-2 mA
5
0.2-2 µA
8
0.2-2 nA
In example, resistance ranges for a voltage of 100 VDC:
N°
Range
N°
Range
N°
Range
N°
Range
0
AUTO
3
500 kΩ – 5 MΩ
6
500 MΩ – 5 GΩ
9
500 GΩ – 5 TΩ
1
5 kΩ – 50 kΩ
4
5 MΩ – 50 MΩ
7
5 GΩ – 50 GΩ
10
5 TΩ – 50 TΩ
2
50 kΩ – 500 kΩ
5
50 MΩ – 500 MΩ
8
50 GΩ – 500 GΩ
Below the min. value the display will indicate a
"UNDER RANGE"
message and - - - -
Beyond the max. value, the display will indicate a
"OVER RANGE"
message and will take
correct measurements up to 10 times the maximum value of the range.