SEFELEC1500M Instructions
145
V1.06
15.1.4.
DEVICE-DEPENDENT COMMANDS
MEG (MEGohmeter)
Activates the megohmeter function. Must be sent from the starting pattern.
PIC (PICOammeter)
Activates the current measurement function. Must be sent from the starting pattern.
SEQ (SEQuence)
Activates the sequence function. Must be sent from the starting pattern.
CONF (CONFig)
Activates the configuration function. Must be sent from the starting pattern.
PAR (PARameter) <NR1>
Selects the parameter set for the active function. E.g.:
MEG:PAR 1
DCV (DCVoltage) <NR1>
Changes the value of the test voltage of the current parameter set for the active func-
tion and switches to DC voltage if the option is present in strength.
E.g.: DCV 500
HTIM (HTIMe) <NR1>
or <NR2>
Changes the value of the maintaining time in seconds of the current parameter set for
the active function. Values between 0.1 and 9,999
E.g.: HTIM 3
RTIM (RTIMe) <NR1> or <NR2>
Changes the value of the rising time in seconds of the current parameter set for the ac-
tive function. Values between 0 and 9,999
E.g.: RTIM 10
FTIM (FTIMe) <NR1> or <NR2>
Changes the value of the falling time in seconds of the current parameter set for the ac-
tive function. Values between 0 and 9,999
E.g.: FTIM 5
HLIM (HLIMIt) <NR3>
Changes the value of the upper threshold of the current parameter set for the active
function.
E.g.:
MEG:HLIM 2.0E+6
(new value in Ohms)
PIC:HLIM 1.45E-6
(new value in Amperes)
The unit is automatically expressed in Volts, Ohms or Amperes and depends on the
function.
CBHLIM <NR3>
Same as HLIM but resetting of the lower threshold before loading the value of the upper
threshold
.