Instructions for use Anesthetic gas receiving system AGS
11
English/English US
Validated reprocessing procedures
Overview of the reprocessing procedures of the
components
Surface disinfection with cleaning
Categories:
– Flow tube
Prerequisites:
– The surface disinfectant has been prepared in
accordance with the manufacturer's instructions.
– The manufacturer's instructions, e.g., regarding
shelf life or application conditions, are observed.
– An uncontaminated, lint-free cloth soaked in sur-
face disinfectant is used for the cleaning surface
disinfection.
Cleaning
1
Wipe off obvious soiling with a disposable cloth
soaked in surface disinfectant. Dispose of the
cloth.
2
Wipe all surfaces. After that, there must no lon-
ger be any soiling visible.
Surface disinfection
3
Wipe cleaned surfaces again to visibly wet all
surfaces to be disinfected with surface disinfec-
tant.
4
Wait for the surface disinfectant contact time.
5
At the end of the contact time, moisten a new,
uncontaminated, lint-free cloth with water (at
least drinking water quality).
6
Wipe all surfaces until no remains of the surface
disinfectant, such as foam residues or streaks,
are visible.
7
Wait until the surfaces are dry.
8
Check the surfaces for visible damage and, if
necessary, replace the product.
Supplementary information
Manual precleaning (AGS header only)
1
Prepare the cleaning solution in accordance with
the manufacturer's instructions.
2
Immerse the product in the solution, making
sure to displace trapped air.
3
At the beginning and towards the end of the con-
tact time, swirl the product backwards and for-
wards several times in the solution. Make sure
that the solution reaches all surfaces and interior
spaces.
4
After the contact time has elapsed, rinse the
product under running water until residual clean-
ing agent is no longer discernible.
5
Check the product for visible damage and re-
place if necessary.
6
Thoroughly shake out residual water.
Refer to the sections "Preparation" and "Opera-
tion", see page 9.
Machine cleaning with thermal
disinfection
Use a washer-disinfector that meets the require-
ments of the standard ISO 15883. Dräger recom-
mends the use of a load carrier for anesthesia ac-
cessories and ventilation accessories. Follow the
manufacturer's instructions for the washer-disinfec-
tor.
Components:
– AGS header
– Container for buffer volume
– Transfer hoses and scavenging hoses
– Plugs and adapters
Prerequisites:
The washer-disinfector has been prepared in accor-
dance with the manufacturer's instructions.
Positioning the components in the load carrier
Procedure:
1
Position the components to be stable.
2
Ensure the following:
– All surfaces and interiors can be completely
rinsed.
– The water can drain off freely.
Performing reprocessing
1
Select a cycle.
2
When the cycle has ended, check the compo-
nents for visible soiling and repeat the cycle if
necessary.
3
Check the components for visible damage and
replace if necessary.
Storage and transport
After reprocessing, there are no special require-
ments for storage and transport of the product. How-
ever, the following must be observed:
– Store dry and free of dust
– Avoid recontamination and damage during
transport
All further information on storage and transport in-
cluded in the accompanying documents must be ob-
served.
Other agents
Disinfectants
Use nationally approved disinfectants suitable for
the respective reprocessing process and the intend-
ed application.
Surface disinfectants
The manufacturers of the surface disinfectants have
verified at least the following spectra of activity:
– Bactericidal
– Yeasticidal
– Virucidal or virucidal against enveloped viruses
Follow the manufacturer's instructions for surface
disinfectants.
The following surface disinfectants were compatible
with the material at the time of testing:
Dräger states that oxygen-releasing agents and
chlorine-releasing agents may cause color change
in some materials. Color change does not indicate
that the product is not functioning correctly.
Category
Surface disin-
fection with
cleaning
Machine clean-
ing with thermal
disinfection
Flow tube
Yes
No
AGS head-
er
No
Yes
Container
for buffer
volume
No
Yes
Transfer
hoses and
scaveng-
ing hoses
No
Yes
Plugs and
adapters
No
Yes
Surface disin-
fectant
Manufac-
turer
Con-
centra-
tion
Contact
time
Dismozon plus
BODE
Chemie
1.6 %
15 min
Neodisher LM2
Dr. Weigert 2 %
10 min
WARNING
Risk due to penetrating liquid
Penetrating liquid may cause the following:
– Damage to the device
– Device malfunctions
Ensure that no liquid penetrates the device.
Step
Agent
Man-
ufac-
turer
Con-
cen-
tra-
tion
Tem-
pera-
ture
Con-
tact
time
Prelimi-
nary
cleaning
Tap wa-
ter
–
–
Tap wa-
ter tem-
perature
Min.
2 min
Cleaning
Neo-
disher
Medi-
clean
forte
Dr.
Wei-
gert
–
–
Min.
5 min
Neutrali-
zing
Neo-
disher
Z
Dr.
Wei-
gert
Min.
0.1
%
Tap wa-
ter tem-
perature
Min.
1 min
Flushing Demi-
nera-
lized
water
–
–
Tap wa-
ter tem-
perature
Min.
1 min
Disinfec-
ting
–
–
–
Min.
93 °C
(199.4 °
F)
Min.
5 min
Drying
–
–
–
–
Drying
time
de-
pends
on the
load
Class of
active in-
gredient
Surface dis-
infectant
Manufac-
turer
Listing
Chlorine-
releasing
agents
BruTab 6S
Brulin
EPA
1)
1) United States Environmental Protection Agency
Clorox Pro-
fessional Dis-
infecting
Bleach
Cleaner
Clorox
EPA
Dispatch
Hospital
Cleaner Dis-
infectant Tow-
els with
Bleach
Klorsept 17
Meden-
tech
EPA
Actichlor plus Ecolab
USA
EPA
Oxygen-re-
leasing
agents
Descogen
Liquid
Antisepti-
ca
CE
Descogen
Liquid r.f.u.
Oxygenon
Liquid r.f.u.
Dismozon
plus
BODE
Chemie
CE
Oxycide
Ecolab
USA
EPA
Perform
Schülke &
Mayr
CE
SteriMax
Wipes
Aseptix
CE
Incidin Oxy-
Wipes
Ecolab
USA
CE
Quaterna-
ry ammoni-
um com-
pounds
acryl-des
2)
2) Virucidal against enveloped viruses
Schülke &
Mayr
CE
Mikrozid alco-
hol free liq-
uid
Mikrozid alco-
hol free
wipes
Mikrozid sen-
sitive liquid
Mikrozid sen-
sitive wipes
Cleanisept
Wipes Maxi
Dr.
Schum-
acher
CE
Surfa'Safe
Premium
ANIOS
Laborato-
ries
CE
Wip'Anios Ex-
cel
Tuffie 5
Vernacare ARTG
3)
3) Australian Register of Therapeutic Goods