5 Care, cleaning and maintenance by the user
5.3 Instruments and instrument hoses
Dentsply Sirona
Operating Instructions Simulation Unit Sim Multi Bench
86
66 82 384 D3684
D3684.201.01.02.02 07.2019
5.3.3
Lubricate, disinfect/sterilize the treatment instruments
5.3.3.1
Treatment instruments
The procedures required for the following treatment instruments are
described in the following separate operating instructions:
● BL motor
● BL ISO motor
● Straight and contra-angle handpieces in various versions
● Turbines
● 3-way syringe Sprayvit
● Standard3-way syringe, see "Cleaning and sterilizing the standard 3-
way syringe"
● SiroSonic L Scaler
● Cavitron scaler
5.3.3.2
Cleaning and sterilizing the standard 3-way syringe
H O
2
134°C 274°F
3 min
The Standard 3-way syringe can be cleaned under running water. Spray
and wipe disinfection can be used. The nozzle can be removed from the
tip for sterilization.
➢ Press on the ring-shaped unlock button.
Ä
The jet adapter is released. The jet will fall out.
To attach the nozzle, press it in until it clicks audibly into place.
WARNING
The nozzle of the Standard 3-way syringe must lock into the holder.
Otherwise, the jet could fall out during usage.
➢ Prior to usage, ensure that the nozzle is securely in place.