4 Operation
4.4 Dentist element
Dentsply Sirona
Operating Instructions Simulation Unit Sim Multi Bench
40
66 82 384 D3684
D3684.201.01.02.02 07.2019
B
ü
With motor removed and the Endo function switched on:
Switch Autoreverse function on/off.
(The Endo function status display is lit.)
A/B
➢ Press and hold the Counterclockwise/user profile key (> 2 s).
/
Ä
The AutoReverse function is switched on/off.
A horizontal bar is displayed above "bur drive calibrated" or "bur
drive not calibrated" on the dentist element SIM display.
ü
With instruments in place and Endo function switched on:
selecting the purge and sanitizing functions
Purge
➢ Press the Purge / Endo key.
B
Ä
The status display Endo is lit.
Ä
The Purge operating context is shown on the EasyPad display.
3
➢ Press favorite key 3 twice.
Ä
The operating context "Sanitization" appears on the SIM display
of the dentist element.
Customer-specific key / function option
#
With the hash key, a freely available relay (230 VAC, 6A) can be
selected and assigned a customer-specific switch or key function.
(connection by the technician)
The hash key can be configured in the setup dialog.
Display mode / clean
This function switches the user interface of the dentist element SIM to
inoperative.
Clean
ü
With instruments in place: Deactivates the entire user interface of
the dentist element
➢ Press the Display mode / Clean fixed key
Ä
The message "C.L.E.A.N" appears on the SIM display of the
dentist element.
This switches the user interface of the dentist element SIM to
inoperative.
The instruments of the dentist element can no longer be
activated. The Sivision monitor and the camera are switched
off.
➢ Press the Display mode / Clean fixed key (> 3 s).
Ä
The user interfaces are now reactivated.