4.7.3
Presetting the timer
The maximum setting for the timer is 9 hours 30 seconds.
ü
The
timer function setting focus is displayed.
1.
Use favorite keys 1 and 3 to set the required time.
Increments:
From 0.05 to 1.00 = 5 s steps
From 1.00 to 3.00 = 10 s steps
From 3.00 to 9.30 = 30 s steps
2.
Confirm with the
2 / Setup favorite key.
When the time or a speed or intensity is displayed on the SIM display of
the dentist element, the timer can be started by pressing the
Fn key on
the dentist element. Pressing the key again stops the timer and resets
the timer setting. It can then be restarted.
4.7.4
Switching the key sound on/off
It can be set whether pressing a key on the dentist element causes an
acoustic signal to sound.
ü
The setting focus
Key tone flashes.
1.
Change the value with the favorite keys 1 or 3.
Ä
If there is a 1 in the setting focus, the key tone is switched on.
2.
Confirm with the
2 / Setup favorite key.
4.7.5
Setting the purging time for the purge function
The purging time for the purge function may be set between 20 and 120
seconds.
ü
The setting focus
purge function flashes.
1.
Use favorite keys 1 and 3 to set the purge time.
2.
Confirm with the
2 / Setup favorite key.
4.7.6
Setting the purging time for the auto-purge function
The purge time for the auto-purge function can be set between 60 and
180 seconds.
ü
The setting focus purge function flashes.
1.
Use favorite keys 1 and 3 to set the purge time.
2.
Confirm with favorite key 2 / Setup.
4.7.7
Switching afterblow on/off
After a foot control pedal has been released, the cooling spray
remaining in the instrument head or in the tip of the instrument can be
automatically blown out by a brief activation of the chip blower.
ü
The setting focus
Afterblow flashes.
1.
Change the value with the favorite keys 1 or 3.
Ä
If there is a 1 in the setting focus, the afterblow is switched on.
2.
Confirm with the
2 / Setup favorite key.