![Dentsply Sirona Sim Multi Bench Operating Instructions Manual Download Page 58](http://html1.mh-extra.com/html/dentsply-sirona/sim-multi-bench/sim-multi-bench_operating-instructions-manual_2483697058.webp)
4 Operation
4.4 Dentist element
Dentsply Sirona
Operating Instructions Simulation Unit Sim Multi Bench
58
66 82 384 D3684
D3684.201.01.02.02 07.2019
4.4.8.4
Setting the speed
Ncm
rpm
When the Endo function is switched on, the SIM display of the dentist
element can be toggled between torque and speed to set the speed.
To differentiate between torque and speed settings, the display shows
bur drive calibrated/not calibrated and AutoReverse function on/off only
for the torque.
Speed is not saved on the favorite keys, only torque.
ü
Torque is displayed on the SIM display of the dentist element.
Clean
1.
Press the
Display mode / Clean key.
Ä
The set speed is displayed on the SIM display of the dentist
element in rpm.
1
2
Setup
3
2.
Set the desired speed for the contra-angle handpiece. Press and
hold the favorite keys 1 or 3 (> 1 s).
Ä
The speed is increased or reduced.
S
0
The speed can also be increased or reduced by moving the 4-way
switch plate of the C+ electronic foot switch up or down.
S
0
Moving the 4-way switch plate of the C+ electronic foot control to the left
also toggles between torque and speed.
4.4.8.5
Switching AutoReverse ON/OFF
If the auto-reverse function is switched on, the burr drive automatically
switches to counterclockwise rotation when the torque set is reached. If
the foot pedal is pressed again, the burr drive switches back to
clockwise rotation.
ü
The electric motor is removed from its holder with the Endo function
switched on.
A/B
➢ Press and hold the
Counterclockwise rotation / User profile key
(> 2 s).
/
Ä
The AutoReverse function is switched on/off. On the SIM
display of the dentist element, a horizontal bar is displayed over
the
Bur drive calibrated or Bur drive not calibrated display.