4 Operation
4.4 Dentist element
Dentsply Sirona
Operating Instructions Simulation Unit Sim Multi Bench
54
66 82 384 D3684
D3684.201.01.02.02 07.2019
4.4.8
Endo function
The speed and torque of the rotating instrument can be set precisely
using the Endo function.
With the Endo function, the speed and torque for up to three work steps
can be saved on the user interface.
CAUTION
If your simulation unit does not have the Endo function, no electronic
torque limitation is available.
Root canal files can easily break during operation without torque
limitation.
➢ Do not perform any endodontic treatments without torque
limitation. Use an endodontic handpiece with mechanical torque
limitation, e.g. SiroNiTi from Dentsply Sirona.
CAUTION
Torque limitation does not always protect against file breakage!
The Endo function maintains the torque values which you have set
with a tolerance safety margin.
Please note: The limiting values stated by the file manufacturers are,
as standard, defined by means of uniform friction over the entire
defined working length. As a rule, this is physically not the practice-
relevant therapy situation.
CAUTION
Root canal files are subject to material fatigue.
Fatigued files can break off.
➢ Use files only for the service life specified by the manufacturer.
CAUTION
Only the Dentsply Sirona Endo 6:1 (SN 6407 and above / July 2010)
and Endo 6 L contra-angle handpieces may be used for the Endo
function.
Instruments from other manufacturers can lead to malfunctions. Third-
party instruments may be improperly calibrated for endodontics.
➢ For endodontics, use only Dentsply Sirona Endo 6:1 (SN 6407
and above / July 2010) or Endo 6 L contra-angle handpieces.