5 Care, cleaning and maintenance by the user
5.1 Basics
Dentsply Sirona
Operating Instructions Simulation Unit Sim Multi Bench
78
66 82 384 D3684
D3684.201.01.02.02 07.2019
134°C 274°F
3 min
Sterilization
Sterilization may be conducted for components that are marked
accordingly. Dentsply Sirona recommends sterilization in a steam
sterilizer (autoclave) according to ISO 13060, Class B (e.g., DAC
PREMIUM / DAC PROFESSIONAL).
The sterilization must be completed with multiple vacuum fractionation
(class B sterilizer). The process parameters can be found in the
engraved characters on the relevant components and the operating
instructions for these. Please also observe the operating instructions for
the sterilizer.
5.1.5
Inspection, maintenance and testing
Unless otherwise specified in this operating manual, test all components
for proper functioning on a regular basis and carry out a visual
inspection for damage and wear. Exchange damaged components if
necessary.