5 Care, cleaning and maintenance by the user
5.3 Instruments and instrument hoses
Dentsply Sirona
Operating Instructions Simulation Unit Sim Multi Bench
84
66 82 384 D3684
D3684.201.01.02.02 07.2019
5.3.2.2
Purging water paths manually
The purge function is optional on the simulation unit.
If this function is not available, the water paths must be purged
manually. The water-bearing instruments and the 3-way syringe must
be purged.
The instruments must be removed individually and held over a
sufficiently large waterproof container (e.g. bucket) for purging.
Purge the water lines of all instruments:
● Before starting work
● At the end of the work day
Preparation
max.
➢ Set all of the instruments to maximum water flow.
The water regulators under the instrument holder must be turned
counterclockwise.
Purging water lines
1.
Hold the 3-way syringe over a sufficiently large waterproof container
and press the water button for at least 20 seconds.
2.
Then put the 3-way syringe back in its holder.
max.
3.
Remove a single instrument and hold it over a sufficiently large
waterproof container.
4.
Activate the spray.
Toggle the switch on the pneumatic foot switch to the right.
Activate the left key on the C+ electronic foot control.