5 Care, cleaning and maintenance by the user
5.3 Instruments and instrument hoses
Dentsply Sirona
Operating Instructions Simulation Unit Sim Multi Bench
82
66 82 384 D3684
D3684.201.01.02.02 07.2019
5.3.2.1
Purge function
Setting the purging time
The purging time of the removed instruments can be set between 20
and 120 seconds in the setup for the simulation unit, see “Setting the
purging time for the purge function” [→ 72].
Switching on the purging function
ü
All instruments are in their holders.
Purge
1.
Press on the
Endo / Purge key.
Ä
The operating context
Purge appears on the SIM display of the
dentist element.
2
Setup
2.
Press the
2 / Setup favorite key.
Purging water lines
ü
On the SIM display of the dentist element, a revolving element is
displayed next to the
Purge display.
1.
Remove all instruments from the dentist element that are to be
purged and hold them over a container, e.g. bucket.
Ä
After an instrument is removed, the purge process starts
automatically after about 5 seconds (except for the 3-way
syringe). The removed instruments are purged with water for
the duration of the set purging time.
CAUTION
Do not purge more than two burs simultaneously.
If the simulation unit features the option of a third burr drive, the water
flow rate is reduced when purging simultaneously.
➢ Do not remove more than two burr drives for purging. Purge the
third burr drive separately afterwards.
2.
Place the removed instruments back in their holders after the
purging process.
3.
Hold the 3-way syringe of the dentist element over the container
and press the water button for at least 20 seconds.
2
Setup
4.
Press the
2 / Setup favorite key to end the purge program.
Ä
The time appears on the SIM display of the dentist element.
Ä
The simulation unit is ready for operation again.