![Dentsply Sirona Sim Multi Bench Operating Instructions Manual Download Page 68](http://html1.mh-extra.com/html/dentsply-sirona/sim-multi-bench/sim-multi-bench_operating-instructions-manual_2483697068.webp)
4 Operation
4.6 Operating light
Dentsply Sirona
Operating Instructions Simulation Unit Sim Multi Bench
68
66 82 384 D3684
D3684.201.01.02.02 07.2019
4.6.3
Setting the brightness, color temperature, and sensor
control of the LEDview Plus
The operating light is always switched on with the programmed
brightness level. It is programmed on the user interface of the dentist
element.
Setting LEDview Plus on the user interface
1.
Press and hold the
Operating light key on the dentist element
(> 2 s).
Ä
The operating context
Brightness is displayed on the SIM
display of the dentist element. The setting focus flashes.
1
2
Setup
3
2.
Use the left and right arrow of the favorites key to adjust the
brightness of the operating light (7 levels). The left arrow key
reduces the brightness, the right arrow key increases the
brightness.
2
Setup
3.
Confirm your setting with the
2 / Setup favorites key.
Ä
The operating context
Color temperature is displayed on the
SIM display of the dentist element. The setting focus flashes.
1
2
Setup
3
4.
Use the left and right arrow of the favorites key to adjust the color
temperature of the operating light (3 levels). The left arrow key
reduces the color temperature, the right arrow key increases the
color temperature.
2
Setup
5.
Confirm your setting with the
2 / Setup favorites key.
Ä
The operating context
Sensor control is displayed on the SIM
display of the dentist element. The setting focus flashes.
1
2
Setup
3
6.
Use the left and right arrow of the favorites key to adjust the
operating distance of the contactless sensor (5 levels). The left
arrow key reduces the operating distance, the right arrow key
increases the operating distance. You can check the setting directly
on the operating light without leaving the settings screen. The
adjustment should be selected in such a way that unintended
operation of the contactless sensor is not possible. Set the
operating distance to "0", in order to deactivate the contactless
sensor.
2
Setup
7.
Confirm your setting with the
2 / Setup favorites key.
Ä
The clock time is displayed.