3 Unit description
3.1 Technical data
Dentsply Sirona
Operating Instructions Simulation Unit Sim Multi Bench
18
66 82 384 D3684
D3684.201.01.02.02 07.2019
3
Unit description
3.1
Technical data
Model designation
Simulation workstation Sim Multi Bench
Power connection
100-240 V AC ± 10%
50/60 Hz
Rated current
3.3 A - 1.5 A at 100 - 240 V
Main building fuse
Type B automatic circuit
breaker
100 - 115 VAC: 20 A medium-blow
220 - 240 VAC: 16 A medium-blow
Overvoltage category
II according to IEC 60664-1
Protection class
I
Media
Water Quality
Drinking water
Water hardness
1.4 mmol/l (=8°dH) -2.1 mmol/l
(=12°dH)
Air quality
oil-free, dry, non-contaminated and free
of dirt
Power ratings
On-site water pressure
2.5-6 bar (≥2.5 l/min)
On-site air pressure
5.5-7.5 bar (≥ 50 l/min)
On-site suction air vacuum
P ≤0.18 bar (≥ 350 l/min)
Waste water
max. 3 l/min
Operating conditions
+10
+40
1060
700
85
30
Intended use
For indoor use
Ambient temperature
10°C-40°C (50°F-104°F)
Relative humidity
30%-85%
Air pressure
500hPa-1060hPa
Installation site
≤3000 m above sea level
The simulation unit is not suitable for
operation in explosion hazard areas.
Degree of contamination
2 acc to. IEC 60664-1
Dimensions
WxHxD (without light)
approx. 1010 cm x 920 cm x 620 cm
WxHxD (with light)
approx. 1010 cm x 2030 cm x 620 cm
Weight