Profi
Message
Nernst factor = ln10 * molar gas constant * absolute temperature / Faraday con-
stant
molar gas constant = 8.314472 in J/(mol*K),absolute temperature = temperature
of the reference ju 273.15 in K,Faraday constant = 96.4851534 in J/(mV*-
mol)
Sensor type
Select
pH
.
Sensor measuring range
A range of ±625 mV is sufficient for this sensor type.
Temp. compensation
Temperature compensation is determined by measuring the temperature at the
measurement point in the solution to be determined. Activate compensation and
configure the channel using the
Advanced settings
Scaling
The default scaling uses the standard measurement
methods and should not be altered.
4.6.3.8
Advanced settings tab
The
Extended settings
tab is identical for all analogue input sensors but not every
setting is accessible from every type of sensor.
89
Delphin Technology AG
Dezember 2015
Profi
Message Manual