Profi
Message
Operation can reduce the number of analogue inputs from 8 to 4/2/1. This
reduces the time for evaluation and for measurement data transmission, or
increases the trigger rate. Some functions are available only for settings of ≤4
channels. Refer also to the section
.
This setting takes priority over the analogue inputs
marked as active.
Sample rate
Sets the sampling frequency in the 256 Hz to 20480 Hz range. The setting is
applied to all selected analogue inputs. The
Operating mode
settings affect the
options that are available here.
Number of samples
The number of samples is set (i.e. measurement values) in the 128 to 8192 range
for a time signal. The setting is applied to all selected analogue inputs. The
Oper-
ating mode
settings affect the options that are available here.
Recording time
Information about the time signal. It is determined from the sampling frequency
and the number of samples and then displayed here.
Frequency range/number of lines/frequency resolution
Information about the frequency spectrum. It is determined from the sampling fre-
quency and number of samples and then displayed here.
Time signal tab
287
Delphin Technology AG
Dezember 2015
Profi
Message Manual
Summary of Contents for ProfiMessage
Page 1: ......
Page 31: ...ProfiMessage 3 5 4 ADIT 31 Delphin Technology AG Dezember 2015 ProfiMessage Manual ...
Page 32: ...ProfiMessage 3 5 5 ADVT 3 5 6 AMDT 32 Delphin Technology AG Dezember 2015 ProfiMessage Manual ...
Page 33: ...ProfiMessage 3 5 7 DIOT 3 5 8 IOIT 33 Delphin Technology AG Dezember 2015 ProfiMessage Manual ...
Page 314: ...ProfiMessage 314 Delphin Technology AG Dezember 2015 ProfiMessage Manual ...
Page 322: ......