Measurement accuracy is determined by acquisition of the fn basic frequency via
the digital frequency input.
Select the settings
Sampling frequency
and
Number of samples
to enable acquis-
ition of more than one but less than ten signal periods in a time signal block. This
achieves an accuracy of ±1
…
5 degrees for the phase angle.
The following table is for a frequency spectrum with 2048 and 4096 lines.
Acquired sig-
nal periods
>1
2
5
10
20
Measurement
deviation
phase angle
±1 degrees ±2 degrees ±3 degrees ±5 degrees ±10 degrees
The rule of thumb for a measurement deviation of ±5 degrees:
Examples
Sample rate
2560 Hz 2560 Hz
1280 Hz
Number of samples
2048
4096
4096
Input signal range for a measurement devi-
ation of ±5 degrees
1.25 to
25 Hz
0.625 to
12.5 Hz
0.3125 to
6.25 Hz
Converting a sine definition to a cosine definition
279