Profi
Message
The
End block value
setting transmits a frequency spectrum of the corresponding
averaging level only after n measurements.
Order analysis
When order analysis is enabled, the evaluation software (e.g. ProfiSignal) will be
informed that the measured frequency spectrum is to be displayed as an order
spectrum. The spectrum is then normalized to the measured rotation speed. This
requires that rotation-pulse is connected and frequency measurement is activ-
ated.
Rotation measurement tab
Counter mode
Both counters are assigned directly to the first two digital inputs.
Two modes are possible:
l
Pulse counting
l
Frequency measurement
Settings are simultaneously effective for both counters.
Pulse counting
is con-
secutive (16-bit = 65536) and reset is not possible.
Frequency measurement
takes place automatically in Hz. More information is available in the section
quency and rotation speed measurement
Rotation speed measurement, source
290
Delphin Technology AG
Dezember 2015
Profi
Message Manual
Summary of Contents for ProfiMessage
Page 1: ......
Page 31: ...ProfiMessage 3 5 4 ADIT 31 Delphin Technology AG Dezember 2015 ProfiMessage Manual ...
Page 32: ...ProfiMessage 3 5 5 ADVT 3 5 6 AMDT 32 Delphin Technology AG Dezember 2015 ProfiMessage Manual ...
Page 33: ...ProfiMessage 3 5 7 DIOT 3 5 8 IOIT 33 Delphin Technology AG Dezember 2015 ProfiMessage Manual ...
Page 314: ...ProfiMessage 314 Delphin Technology AG Dezember 2015 ProfiMessage Manual ...
Page 322: ......