Frequencies can be measured between the lower limit at 2*fs/ns and the upper
limit at fs/10 for selected parameters (recording parameters). The measurement
value equals 0 beyond these limits.
Below is the corresponding configuration dialogue (
Main settings
tab):
9.2.5.11
"Maximum vect. sum" characteristic value
The DSP performs vectoral addition of all amplitude values from the time signals of
two consecutive channels (e.g. analogue input 1 and 2) and then determines the
maximum. The characteristic value must be set for the first of the two consecutive
channels.
Options
Smooth curve
/
Discard anomalies
Smooth curves prior to determining characteristic values may be applicable to peri-
odic signals that have noise or distortions. These two settings are available in the
Options
tab:
9.2.5.12
"Artihm. average of product”
The DSP performs multiplication of all amplitude values from the time signals of
two consecutive channels (e.g. analogue input 1 and 2) and then determines the
arithmetical average. The characteristic value must be set for the first of the two
consecutive channels.
269
Delphin Technology AG
Dezember 2015
Profi
Message Manual
Summary of Contents for ProfiMessage
Page 1: ......
Page 31: ...ProfiMessage 3 5 4 ADIT 31 Delphin Technology AG Dezember 2015 ProfiMessage Manual ...
Page 32: ...ProfiMessage 3 5 5 ADVT 3 5 6 AMDT 32 Delphin Technology AG Dezember 2015 ProfiMessage Manual ...
Page 33: ...ProfiMessage 3 5 7 DIOT 3 5 8 IOIT 33 Delphin Technology AG Dezember 2015 ProfiMessage Manual ...
Page 314: ...ProfiMessage 314 Delphin Technology AG Dezember 2015 ProfiMessage Manual ...
Page 322: ......