9.2.5.7
"Phase of main vibration" characteristic value
Phase angle measurement is useful only for low fre-
quencies/rotation speeds. The evaluation procedure
(see below) shows why. Further information on phase
angle definition and on measurement precision etc. is
available under
This characteristic value can be determined only when
a cyclical digital pulse is connected to a digital input/-
counter (e.g. the square wave pulse of a rotating
shaft). This signal delivers a stable base frequency fn
as well as a phase reference or trigger signal.
The DSP normalizes the time signal (internally, hidden) to an integral multiple of
the basic frequency fn.
Using the frequency spectrum associated with the line index fn, the DSP determ-
ines the line/vibration component's phase angle 0 … 360°.
9.2.5.8
"Amplitude of the 1X/2X/3X/xX vibration" characteristic value
This characteristic value can be determined only when
a cyclical digital pulse is connected to a digital input/-
counter. This signal delivers the stable basic frequency
fn.
265
Delphin Technology AG
Dezember 2015
Profi
Message Manual
Summary of Contents for ProfiMessage
Page 1: ......
Page 31: ...ProfiMessage 3 5 4 ADIT 31 Delphin Technology AG Dezember 2015 ProfiMessage Manual ...
Page 32: ...ProfiMessage 3 5 5 ADVT 3 5 6 AMDT 32 Delphin Technology AG Dezember 2015 ProfiMessage Manual ...
Page 33: ...ProfiMessage 3 5 7 DIOT 3 5 8 IOIT 33 Delphin Technology AG Dezember 2015 ProfiMessage Manual ...
Page 314: ...ProfiMessage 314 Delphin Technology AG Dezember 2015 ProfiMessage Manual ...
Page 322: ......