Digital inputs are used as trigger sources with simultaneous frequency or rotation
speed (for rotating machinery) measurement and phasing (reference) acquisition
are simultaneously carried out.
9.1.5.6
Digital outputs
Properties
4 digital outputs. Galvanic isolation up to 2.5 kV. Switching voltage at max.
50V DC. Switching current max. 2.5A DC. Protector diode integrated. Flyback
diode external.
Refresh rate
Determined by source channel (e.g. timer channel). Typical range 0 … 10 Hz;
dependent on DSP load.
Application
The digital outputs report warnings or alarms (e.g. generated via limit values for
vibration characteristic values). Signals are also delivered for process control.
9.1.6
Appendix
9.1.6.1
AC coupling for analogue inputs
An analogue input's DC coupling can be switched to AC coupling via two DIP
switches on the signal conditioning board.