AUSTROMAT
®
624 / 644
9.
Appendix
Operating Instructions (Version 01/2012)
143
9.2. Vacuum
Table
The values in the following table were measured at a surrounding pressure of 969 hPa.
1 hPa = 1 mbar = 0.015 psi
1 psi = 69 mbar = 69 hPa
Entry for
Relative pressure
1)
hPa/psi
Absolute pressure
2)
hPa/psi
Easy/professional/p
ressing programs
Classic programs
0% V0
0
0
969
14.0
10
%
V1
52 0.75 917 13.3
20
%
V2
153 2.2 816 11.8
30
%
V3
264 3.8 705 10.2
40
%
V4
365 5.3 604 8.7
50
%
V5
472 6.8 497 7.2
60
%
V6
577 8.3 392 5.7
70
%
V7
685 9.9 284 4.1
80
%
V8
793 11.5 176 2.5
90 %
V9 minimum (=
switching point)
866 12.5 103 1.5
100 %
V9 maximum (=45 s
after V9 minimum)
> 934
13.5
< 35
0.51
Tab. 9-2
Pressures of the different vacuum levels
1) Difference between surrounding pressure and absolute pressure
2) Pressure in the firing chamber
Summary of Contents for AUSTROMAT 624
Page 1: ...Operating Instructions AUSTROMAT 624 644 Version 01 2012 ...
Page 8: ...Table Of Contents AUSTROMAT 624 644 8 Operating Instructions Version 01 2012 ...
Page 146: ...9 Appendix AUSTROMAT 624 644 146 Operating Instructions Version 01 2012 ...
Page 147: ...AUSTROMAT 624 644 9 Appendix Operating Instructions Version 01 2012 147 ...
Page 148: ...9 Appendix AUSTROMAT 624 644 148 Operating Instructions Version 01 2012 ...
Page 149: ...AUSTROMAT 624 644 9 Appendix Operating Instructions Version 01 2012 149 ...
Page 150: ...9 Appendix AUSTROMAT 624 644 150 Operating Instructions Version 01 2012 ...
Page 151: ...AUSTROMAT 624 644 9 Appendix Operating Instructions Version 01 2012 151 ...
Page 152: ...9 Appendix AUSTROMAT 624 644 152 Operating Instructions Version 01 2012 ...
Page 155: ...AUSTROMAT 624 644 10 Index Operating Instructions Version 01 2012 155 ...