5. User interface
Operating Instructions (Version 01/2012)
95
5.7.3.3. Temperature calibration
NOTE:
When calibrating the temperature, observe the following:
•
Only use the DEKEMA calibration set for this type of unit,
•
Preheat the firing oven well,
•
Always use the fiber firing table.
In the "Setup" menu, click the [Temperature calibration] button. The following temperature
calibration window appears. The set CCF value is shown.
Fig. 5-31
Setup > Temperature calibration
Buttons
•
[Start]:
Starts temperature calibration.
•
[ESC]:
Closes the window and switches back to the "Setup" menu. The
changes will not be saved.
Calibrate temperature The procedure for calibrating the temperature is in section 7.2. You
can terminate the program at any time when you click the [Stop]
button. A safety query appears and asks if you really want to end the
program:
•
[OK]: Confirm your decision and end the program.
•
[ESC]: Change your decision and continue the program.
Change CCF value
•
Click on the CCF value. The numeric keyboard window
appears.
•
Enter the desired CCF value.
Summary of Contents for AUSTROMAT 624
Page 1: ...Operating Instructions AUSTROMAT 624 644 Version 01 2012 ...
Page 8: ...Table Of Contents AUSTROMAT 624 644 8 Operating Instructions Version 01 2012 ...
Page 146: ...9 Appendix AUSTROMAT 624 644 146 Operating Instructions Version 01 2012 ...
Page 147: ...AUSTROMAT 624 644 9 Appendix Operating Instructions Version 01 2012 147 ...
Page 148: ...9 Appendix AUSTROMAT 624 644 148 Operating Instructions Version 01 2012 ...
Page 149: ...AUSTROMAT 624 644 9 Appendix Operating Instructions Version 01 2012 149 ...
Page 150: ...9 Appendix AUSTROMAT 624 644 150 Operating Instructions Version 01 2012 ...
Page 151: ...AUSTROMAT 624 644 9 Appendix Operating Instructions Version 01 2012 151 ...
Page 152: ...9 Appendix AUSTROMAT 624 644 152 Operating Instructions Version 01 2012 ...
Page 155: ...AUSTROMAT 624 644 10 Index Operating Instructions Version 01 2012 155 ...