5.7 Infrared communication specifications
By performing infrared communication between the infrared communication area (IrDA) of the
detector and infrared data communication unit connected to the Image Capture Computer and
by enacting the link (detector recognition), the initial settings to be established when using the
CXDI-80C Wireless can be selected and the settings used between a multiple number of
imaging rooms can be switched. The table below lists the infrared communication
specifications.
Item
Details
Remarks
Timeout time
3 seconds
Time when infrared
communication is tried
after the power switch
of the detector has been
pressed.
Communication
environment
Communication
range
Within 30 cm
Environment
conditions
•
This form of communication cannot be
used in an environment where infrared
signals are blocked.