1
1
.
.
O
O
v
v
e
e
r
r
v
v
i
i
e
e
w
w
- 9 -
5.2 Power supply specifications
Item
Details
Remarks
Battery charger
Battery charger
AC 100 to 240 V,
50/60 Hz, 70 to 90 VA
Power cord set
X-ray interface box
X-ray interface box
DC 12V, 0.25 A
AC adapter
AC 100 to 240 V,
47/63 Hz, 0.33 to
0.18 A
Power cord set
Wiring unit (option)
AC/DC box
AC 100 to 240 V,
50/60 Hz, 0.29 to
0.18 A, 29 to 41 VA
Power cord set
5.3 Recommended LAN cable specifications
The LAN used by the CXDI system shall be a local network, and it shall be isolated from the
clinic/hospital network. etc. The specifications of the LAN cables are given below.
CAT-5 or above
Recommended cable length: Max. 30 meters
Straight type (type may differ depending on the hub functions)
Item
Details
Remarks
For connecting PC and hub
1000Base-T
For connecting X-ray interface box and hub
100Base-Tx
For connecting wireless LAN access point and hub
1000Base-T
For connecting wiring unit and hub
100Base-Tx
5.4 Environmental resistance specifications
Item
Operation and
installation conditions
Transportation and
storage conditions*
Remarks
Temperature
range
+5 to +35 [°C]
-30 to +50 [°C]
Use in an ambient temperature
exceeding 35°C is not
considered as constituting
normal use.
Humidity
range
30 to 80 [%RH]
10 to 95 [%RH]
No condensation
Vibration
0.05 [G], 10 to 100 [Hz]
1.0 [G], 10 to 100 [Hz]
Atmospheric
pressure
700 to 1060 [hPa]
700 to 1060 [hPa]
* The storage environment specifications apply in the formally packaged status. The
specifications in the non-formally packaged status (including no packaging) are +5 to
+40 [°C] for the temperature range and 30 to 85 [%RH] for the humidity range.
Summary of Contents for CXDI-80C
Page 6: ...Note These measurements are only guidelines ...
Page 10: ...1 1 O Ov ve er rv vi ie ew w 2 2 System diagram ...
Page 48: ...2 Installation procedures 15 4 Handle unit option Corner pads Outer box Inner box ...
Page 133: ...Rev 01 CONTENTS Detector unit 1 X ray interface box unit 2 Wiring unit 4 Handle unit 6 ...