4. Repair Guide
- 12 -
3.3.5.2 Mounting the AC/DC power supply
1) Slide the AC/DC power supply in the direction of the red arrow shown in the figure, and
insert the switch part into the hole in the base.
2) While checking the clearance between the hole in the base and area around the switch part,
adjust the position of the AC/DC power supply so that the clearance is uniform on the left and
right, and loosely secure it using the five screws (XB1-2300-606).
3) Attach the two screws (XB1-2300-606) on the front panel of the AC/DC box used to secure
the AC/DC power supply.
4) Tighten up the five screws (XB1-2300-606) that were loosely secured.
Reference:
Tightening torque: 58.8 N-cm (6 kgf-cm)
XB1-2300-606 screws (x 5)
Loosely tighten these screws.
AC/DC
power supply
Insert the switch part into
the hole in the base.
Press the AC/DC power supply
against the inside of the base.
AC/DC box base
Adjust the position of the AC/DC power supply so that
the clearance is uniform on the left and right.
XB1-2300-606 screws (x 2)
Tighten up only the screws on the
far left and far right.