5.11 [Supplementary notes] Long-term storage of batteries
The characteristics of lithium-ion batteries are such that when the batteries are left standing in
storage for a prolonged period of time, their charging capacity decreases so they may end up
becoming unusable. The higher the temperature in the storage location, the faster the charging
capacity deteriorates. When the batteries become overdischarged, this deterioration progresses
at an even more significant pace.
Storage characteristics when the batteries are charged 40%
Storage
temperature
Time taken to reach
overdischarge region
Time taken to reach
deep discharge region
60
℃
0.7 year
1.2 years
40
℃
0.9 year
1.8 years
25
℃
1.2 years
2.2 years
0
℃
1.3 years
2.3 years
Storage characteristics when the batteries are charged 100%
Storage
temperature
Time taken to reach
overdischarge region
Time taken to reach
deep discharge region
60
℃
1.2 years
1.7 years
40
℃
1.9 years
2.7 years
25
℃
2.3 years
3.3 years
0
℃
2.5 years
3.5 years
* Based on the above characteristics, the batteries may become unusable when they have been
left standing in a storage temperature of 25 to 40°C for more than 1.5 years or so when
charged 40% or for more than 2.5 years or so when charged 100%.
Capacity deterioration characteristics after storage for 1 year
Storage temperature
40% charge
100% charge
60
℃
-22% -39%
40
℃
-10% -16%
25
℃
-3% -10%
0
℃
-1% -5%
* It is not possible to restore the capacity of batteries which has deteriorated as a result of
storage. After storage, the charging capacity will be reduced.
* When batteries with a 100% charge are left standing in a 25°C environment for one year,
their charging capacity will drop by about 10%.
* “100% charge” is the status in which charging with the battery charger is completed.
* Concerning battery control undertaken by the sales companies (or agents)
The batteries have a 100% charge when they are shipped from the manufacturing plant, and
the shipment date is entered on the packing labels. When batteries have been kept in the
inventory of the sales companies (or agents) for more than one year or when two years have
elapsed from the shipment date on the packing labels, the batteries must be 100% charged,
and the sales companies (or agents) must take responsibility for battery control.
Overdischarge region: Cell voltage of 2.5 V or below
Deep discharge region: Cell voltage of 1.3 V or
below, no further charging possible
Summary of Contents for CXDI-80C
Page 6: ...Note These measurements are only guidelines ...
Page 10: ...1 1 O Ov ve er rv vi ie ew w 2 2 System diagram ...
Page 48: ...2 Installation procedures 15 4 Handle unit option Corner pads Outer box Inner box ...
Page 133: ...Rev 01 CONTENTS Detector unit 1 X ray interface box unit 2 Wiring unit 4 Handle unit 6 ...